This walkthrough covers the complete end-to-end flow for a multi-plate assay experiment in LKB Biologics. By the end, you'll have created plate templates, set up an assay design, plated samples, imported instrument data, reviewed results, and carried forward your best hits.
Prerequisites: You need Editor permissions in your project and at least one sample type with samples loaded.
1. Set up Plate Templates ← define your well layouts (controls, replicates, zones)
2. Set up an Assay Design ← define what you're measuring; must enable Plate Metadata
3. Create a Plate Set ← from your sample table; assigns samples to plates
4. Verify plate layout ← use the Plate Viewer to confirm well assignments
5. Import assay data ← drag-and-drop your instrument file
6. Review results & mark hits ← filter, analyze, mark candidates
7. Pick hits to a new plate set ← carry forward for the next round
Steps 1 and 2 are one-time setup. Once templates and assay designs exist, you start from step 3 each experiment.
Plate templates define the layout of your plates — where samples, controls, standards, and replicates go — without any specific sample data. Set these up once and reuse them across experiments.
If you're running a multi-plate experiment with different layouts per plate (e.g., four plates with different replicate patterns), create a separate template for each.
Tip: Template well groups directly affect how results aggregate in charts. If replicates aren't grouped correctly in the template, they may not roll up as expected in your results. Double-check your replicate group assignments before saving.
An assay design tells LKB what data your instrument produces and how to structure it.
Important: The "Plate Metadata" checkbox is easy to miss and cannot be added to an existing assay design after creation. If you forget it, you'll need to create a new assay design.
Critical: The plate names you enter in LKB must match the plate identifiers in your instrument output file exactly. If they don't match, import will fail or data will be misaligned.
Before importing data, confirm your plates look right.
If anything looks wrong, you can edit well data directly in the Grid View (as long as no assay data has been imported yet).
On success, LKB links your instrument measurements to the samples in each well. You're taken to the Plate Set Assays page where results are ready to review.
Supported file formats:
Note on reimport: Reimporting data for a plate is not additive — it clears existing data for that plate and replaces it. This allows you to correct errors without creating duplicates.
After import, results appear on the Assays tab of your plate set.
For routine assays, you can set automatic hit selection criteria in your assay design so that results are flagged on import without manual review. See the reference doc for details.
About replicates in results: If you set up replicate groups in your template, LKB calculates mean and standard deviation across replicate wells automatically. These aggregated values are available as columns in your results and can be used as hit selection criteria. If you're missing replicates in your analysis, check that the well groups in your template are correctly defined.
Why do I only see some of my replicates in charts or results?
Check the replicate group assignments in your plate template. Wells must be assigned to the same group to be treated as replicates. Also confirm none of the replicate wells are empty or QC-excluded.
Why did import fail saying the plate wasn't found?
The plate names in LKB must exactly match the plate identifiers in your instrument file (e.g., the section headers in your Excel output). Edit the plate names in the plate set detail page so they match.
I forgot to check "Plate Metadata" in my assay design. What do I do?
You'll need to create a new assay design with Plate Metadata enabled. This option cannot be added to an existing design after creation.
Can I add more samples to a plate set later?
Yes, if the plate set is eligible (not archived, no child plate sets). Select the additional samples from the sample grid and use Plates → Add to Plate Set, then choose the existing set. New samples are placed starting from the first available well after existing samples.
What happens when I reimport assay data?
Reimport clears existing results for the targeted plates and replaces them — it is not additive. Other plates in the plate set are unaffected.
Can I edit plate well assignments after running an assay?
No. Once an assay run is associated with a plate, that plate becomes read-only. You can reimport the assay data to update results, but the sample-well layout itself cannot be changed.